| Model Number |
DDBB1D4 |
| Medical Device Problem Codes |
Inappropriate/Inadequate Shock/Stimulation (1574); Device Sensing Problem (2917)
|
| Health Effect - Clinical Code |
Shock from Patient Lead(s) (3162)
|
| Date of Event |
03/14/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continuation of d10: 1882tc46 lead, implanted: (b)(6) 2013.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
It was reported that the implantable cardioverter defibrillator (icd) detected ventricular fibrillation (vf) and delivered a shock.It was suspected that the rhythm was atrial fibrillation (af) with rapid ventricular response (rvr), therefore the shock was inappropriate.Additionally, it was noted that there was undersensing on the right ventricular (rv) lead prior to detection that did not appear to delay detection.The icd and lead remains in use.No further patient complications have been reported as a result of this event.
|
| |
|