Catalog Number 46490
Medical Device Problem Code Crack (1135)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the reported device was not returned, angiodynamics is unable to perform a device evaluation.However, a picture was provided showing the purple hub is cracked.The customer's reported complaint description of picc hub cracked was confirmed with the photo provided by the customer.Without receiving product for evaluation, we are unable to definitively determine a root cause for this incident.The most likely root cause for the cracked female luer is due to an over tightened connection with a mating male luer (likely a needleless connector).A contributing factor to the over-tightened connection may be the mechanics/hand placement during infusion access.It is preferable to grasp the needleless connector (nc) when connecting a syringe/tubing to it rather than grasping the luer hub while making a connection with the nc.Grasping the luer hub while connecting a syringe/tubing set to the nc can transmit additional torque to the female luer hub.Inadequate flushing of the device lumen may also be a contributing factor.A review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.Labeling review: the dfu that is supplied in bioflo kits item number (16600281-01) contains the following precaution: the bioflo* midline catheter with endexo* technology is a radiopaque, polyurethane catheter with luer lock hub(s), polyurethane extension tube(s) and suture wing.The catheter is available in single and dual lumen configurations.The bioflo midline is clearly labeled on all available catheter surfaces to identify as a midline versus a traditional picc.Precaution: if a needleless connector is attached to catheter hub, first ensure that it will sustain power injection.Attach power injector to selected lumen hub per manufacturer's recommendations, and open clamp.Stabilize catheter and luer lock hub during dressing removal to prevent accidental dislodgment.Separate dressing away from luer lock hub and toward insertion site.As you separate, keep any tape and dressing close to patient's arm to avoid dislodging catheter or sutures.To remove catheter, grasp catheter between suture wing and insertion site and remove slowly, in small increments, keeping catheter parallel to skin surface.Do not grasp luer lock hub to remove catheter, as catheter damage may occur.It is recommended that only luer lock accessories be used with the bioflo midline catheter with endexo technology.Repeated over-tightening may reduce hub connector life.Do not use hemostats to secure or remove devices with luer lock hub connections.A review of the angiodynamics complaint system noted no adverse trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Reference (b)(4).
 
Event or Problem Description
An end user reported an issue with a midline catheter from a bioflo midline 5f dl-20cm max sterile barrier drape kit w/ two nitinol guidewire.Nine days post procedure, it was reported the patient's midline had cracked on the hub.The midline catheter was removed and replaced.The patient did not experience any adverse effects or harm because of this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIOFLO MIDLINE
Common Device Name
MIDLINE CATHETER
Manufacturer (Section D)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer (Section G)
ANGIODYNAMICS
10 glens falls technical park
glens falls NY 12801
Manufacturer Contact
alexandra invencio
26 forest street
marlborough, MA 01752
5086587990
MDR Report Key19201830
Report Number1317056-2024-00070
Device Sequence Number12744576
Product Code FOZ
UDI-Device IdentifierH965464901
UDI-PublicH965464901
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K161866
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Health Professional,User Facility
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 04/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/31/2025
Device Catalogue Number46490
Device Lot Number5812025
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/05/2024
Initial Report FDA Received Date04/29/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/02/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-