Catalog Number AEHA35
Medical Device Problem Code Device Damaged by Another Device (2915)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Type of Reportable Event Serious Injury
Event or Problem Description
According to the available information the hydrophilic tip appeared to have sheared off the wire.It is believed that the laser fibers came in contact with the hydrophilic tip of the wire and frayed it slightly during the procedure.The tip was not seen via x-ray.
 
Additional Manufacturer Narrative
No other complaints were found for the lot number.
 
Additional Manufacturer Narrative
After receiving this complaint, we searched for other complaint and we didn't find other complaint on the lot number 9413737.Checking the quality databases revealed no anomaly in connection with the described defect.The review of the related production documents did not indicate any deviations or irregularities.All process steps and tests were carried out in accordance with the valid dmr.Based on the event description, we cannot find any indication of a manufacturing defect.The guide wire was checked before use and no damage was found.During the procedure, the laser beams came into contact with the hydrophilic tip of the wire and frayed it slightly.We assume that the laser used may have damaged the guide wire.Since the defective product is not available for investigation, a more detailed root cause analysis cannot be performed.Since no manufacturing defects could be proven, we can only take note of the complaint for statistical purposes and consider it to be unjustified.
 
Event or Problem Description
According to the available information the hydrophilic tip appeared to have sheared off the wire.It is believed that the laser fibers came in contact with the hydrophilic tip of the wire and frayed it slightly during the procedure.The tip was not seen via x-ray.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SOPRANO HYBRID GUIDEWIRE
Common Device Name
GASTRO-UROLOGICAL GUIDEWIRE. SINGLE USE
Manufacturer (Section D)
COLOPLAST A/S
1 holtedam humlebaek, dk 3050
humlebaek 3050
DA  3050
Manufacturer Contact
usbes brian schmidt
1601 west river road n
minneapolis, MN 55411
MDR Report Key19201875
Report Number9610711-2024-00098
Device Sequence Number18154780
Product Code OCY
UDI-Device Identifier03600040326725
UDI-Public03600040326725
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/08/2024
Device Catalogue NumberAEHA35
Device Lot Number9413737_AEHA354002
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/29/2024
Supplement Report FDA Received Date10/02/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/11/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;