It was reported that the tissue pad came off the sonicbeat 5 mm, 35 cm, inline grip at about thirty minutes of use.The pad did not fall into the patient¿s body.The issue occurred during a therapeutic hernia surgery.The procedure was completed with a similar device.There were no reports of patient harm.
|
|
This report is being supplemented to provide additional information based on the device evaluation and legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device was returned for evaluation and the customer's reported issue was confirmed.Based on the results of this investigation, the root cause could not be conclusively determined.It is likely the tissue pad peeled off when the grasping section was closed without grasping tissue and the ultrasound was activated (this includes after tissue was cut).As a result, the tissue pad became worn and subsequently peeled.The event can be detected and prevented by handling the device in accordance with the following instructions for use (ifu): "·do not activate output in seal & cut mode while the grasping section is closed without contacting tissue or vessel, or ensuring that tissue is transected.Otherwise, a local increase of the temperature due to a friction between the probe tip and the grasping section may result in various forms of damage in the probe tip and/or the tissue pad, such as premature wear, breakage, deformation, and/or falling off inside the body cavity and/or partial separating.When cutting and vessel sealing is performed in seal & cut mode, apply light tension on the tissue so that users can confirm it is transected.Also, stop activation immediately after tissue is transected.Otherwise, the grasping section, the tissue pad, or the probe tip may break and fall off, and partial separating of the tissue pad may occur due to a local increase of temperature caused by the friction between tissue pad and the probe tip during activation".Olympus will continue to monitor field performance for this device.
|