Catalog Number IAP-0700 |
Device Problem
Defective Device (2588)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 04/04/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).
|
|
Event Description
|
It wa reported " highly pixilated ap waveform, inaccurate triggering during or case".At the time of this report, the customer has not returned our requests for additional information.If further information is received, the complaint file will be updated.
|
|
Event Description
|
It wa reported " highly pixilated ap waveform, inaccurate triggering during or case".At the time of this report, the customer has not returned our requests for additional information.If further information is received, the complaint file will be updated.
|
|
Manufacturer Narrative
|
Qn#(b)(4).The sample was not returned for investigation; however, the customer provided a video for evaluation.The video shows artifacts on the arterial pressure waveform and missed beats resulting in a lower heart rate which is consistent with the event details.The reported complaint of "highly pixilated ap waveform, inaccurate triggering" is confirmed based on the video provided in the complaint.The video shows artifacts on the arterial pressure waveform and missed beats resulting in a lower heart rate which is consistent with the event details.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the abnormal ap waveform.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
|
|