Catalog Number IAP-0700
Device Problem Defective Device (2588)
Patient Problem Insufficient Information (4580)
Event Date 04/04/2024
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It wa reported " highly pixilated ap waveform, inaccurate triggering during or case".At the time of this report, the customer has not returned our requests for additional information.If further information is received, the complaint file will be updated.
 
Event Description
It wa reported " highly pixilated ap waveform, inaccurate triggering during or case".At the time of this report, the customer has not returned our requests for additional information.If further information is received, the complaint file will be updated.
 
Manufacturer Narrative
Qn#(b)(4).The sample was not returned for investigation; however, the customer provided a video for evaluation.The video shows artifacts on the arterial pressure waveform and missed beats resulting in a lower heart rate which is consistent with the event details.The reported complaint of "highly pixilated ap waveform, inaccurate triggering" is confirmed based on the video provided in the complaint.The video shows artifacts on the arterial pressure waveform and missed beats resulting in a lower heart rate which is consistent with the event details.Based on a review of the device history record (dhr), the product met specification upon release; however, the specifications were not met during the complaint investigation due to the abnormal ap waveform.The root cause of the complaint is undetermined.No further action required at this time.The reported complaint will be monitored for any developing trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL LLC
morrisville NC
Manufacturer (Section G)
ARROW INTERNATIONAL INC.
16 elizabeth drive
chelmsford MA 01824
Manufacturer Contact
katharine tarpley
3015 carrington mill blvd
morrisville, NC 27560
MDR Report Key19202582
MDR Text Key341468173
Report Number3010532612-2024-00317
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902084966
UDI-Public30801902084966
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberIAP-0700
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/04/2024
Initial Date FDA Received04/29/2024
Supplement Dates Manufacturer Received05/24/2024
Supplement Dates FDA Received05/24/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-