• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS TSH; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS TSH; RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE Back to Search Results
Model Number TSH
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
1.Summary of investigation results: the investigation is ongoing.2.Summary of follow up/corrective actions taken: the investigation is ongoing.3.Device returned to manufacturer? no 4.Device labeled "for single use?"? no 5.Device reprocessed and reused? no.
 
Event or Problem Description
1.Describe the event: there was an allegation of questionable high elecsys tsh results for 2 patient samples.2.Patient age range: 1 patient: 86 years, other: asku 3.Patient weight range: asku 4.Patient sex breakdown: 1 f, 1 asku 5.Patient race breakdown: asku 6.Patient ethnicity breakdown: asku.
 
Additional Manufacturer Narrative
Summary of investigation results: no interfering factors were identified.The investigation did not identify a product problem.The cause of the event could not be determined.Summary of follow up/corrective actions taken: one patient sample was submitted for investigation.Device returned to manufacturer? no devices were returned.Device labeled "for single use?"? this device is not labeled for single use.Device reprocessed and reused? this device is not reprocessed or reused.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS TSH
Common Device Name
RADIOIMMUNOASSAY, THYROID-STIMULATING HORMONE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhofer strasse 116
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
amy nelson
9115 hague road
indianapolis, IN 46250-0457
3175216208
MDR Report Key19203013
Report Number1823260-2024-01306
Device Sequence Number8142245
Product Code JLW
UDI-Device Identifier07613336157633
UDI-Public07613336157633
Combination Product (Y/N)Y
Number of Events Summarized1
Summary Report (Y/N)Y
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberTSH
Device Catalogue Number08443432190
Device Lot Number76836101
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by Manufacturer06/21/2024
Initial Report FDA Received Date04/29/2024
Supplement Report FDA Received Date07/02/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient SexUnknown
-
-