| Model Number |
45038 |
| Medical Device Problem Code |
Defective Device (2588)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/01/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3: date of event: used the first date of the month of the aware date as no event date was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure could not be completed.An angiojet console was selected for use.However, it was noted that the roller pump refused to work.Due to this issue, the procedure could not be completed.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event: used the first date of the month of the aware date as no event date was provided.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure could not be completed.An angiojet console was selected for use.However, it was noted that the roller pump refused to work.Due to this issue, the procedure could not be completed.No patient complications were reported.It was further reported that the device kept giving a roller pump error and would not load the catheter.The procedure was completed with a non-bsc device.
|
| |
|
Additional Manufacturer Narrative
|
|
B3: date of event: used the first date of the month of the aware date as no event date was provided.Device evaluated by mfr: the angiojet ultra 5000a was returned for analysis.The work order confirmed that the drawer assembly contributed to the reported event and after fse repair/replacement was complete the console functioned as intended.
|
| |
|
Event or Problem Description
|
|
It was reported that the procedure could not be completed.An angiojet console was selected for use.However, it was noted that the roller pump refused to work.Due to this issue, the procedure could not be completed.No patient complications were reported.It was further reported that the device kept giving a roller pump error and would not load the catheter.The procedure was completed with a non-bsc device.
|
| |
|
Search Alerts/Recalls
|