MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE CAGE GLENOID S, POST AUG, RIGHT; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE CAGE GLENOID S, POST AUG, RIGHT; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 314-13-32
Medical Device Problem Code Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 03/05/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that this female patient's right shoulder was revised 5 years 10 months post op initial surgery.Surgeon voiced his concerns about the glenoid to (b)(6) after a anatomic to reverse revision surgery and after he found out about a recall.Glenoid was revised to competitors device.Patient was last known to be in stable condition following the event.Devices are not returning, hospital kept.
 
Additional Manufacturer Narrative
Section h10: (h3) pending evaluation: (d10) concomitant device(s): ab0646 - 13a2101 - cemex system fast genta 70g, ab0991 - 13a2111 - cemex system fast genta 40g, 1896962 - 300-01-07 - equinoxe, humeral stem primary, press fit 7mm, 5093274 - 300-01-09 - equinoxe, humeral stem primary, press fit 9mm, 5128477 300-20-02 - equinox square torque define screw drive kit, 5182909 - 300-20-02 - equinox square torque define screw drive kit, 5151345 - 300-50-45 - 4.5mm short rep plate, 5028015 - 310-01-47 - equinoxe, humeral head short, 47mm (beta), 4250976 - 314-13-32 - equinoxe cage glenoid s, post aug, right, 8012017045 - a10012 - gps implant kit v2, ab0898r - cnl-09 - small cement cannula.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay corrected information.The following sections were corrected: d2a, d2b, g1, h6.The reason for the revision reported cannot be confirmed from the information provided but may be due to inclusion of the polyethylene in the packaging recall wear, loosening, infection, fracture, instability, dislocation, rotator cuff tear, and/or patient related factors.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and relevant clinical information was not provided.The radiograph provided appears unremarkable.H6: corrected health effect, medical device, component, and investigation clinical codes.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The following sections were updated/corrected: h6.Mdr section codes updated/corrected: b, c, d.The reason for the revision reported cannot be confirmed from the information provided but may be due to inclusion of the polyethylene in the packaging recall wear, loosening, infection, fracture, instability, dislocation, rotator cuff tear, and/or patient related factors.Potential contributions of manufacturing, user, and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and relevant clinical information was not provided.The radiograph provided appears unremarkable.Should additional relevant information be obtained, a follow-up mdr will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE CAGE GLENOID S, POST AUG, RIGHT
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19203914
Report Number1038671-2024-01005
Device Sequence Number20814941
Product Code KWS
UDI-Device Identifier10885862201003
UDI-Public10885862201003
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/31/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/24/2021
Device Catalogue Number314-13-32
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2024
Supplement Date Received by Manufacturer08/11/2025
10/30/2025
Initial Report FDA Received Date04/29/2024
Supplement Report FDA Received Date08/27/2025
10/31/2025
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/27/2016
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-