Model Number RONYX30030UX
Medical Device Problem Codes Burst Container or Vessel (1074); Inflation Problem (1310); Leak/Splash (1354); Improper or Incorrect Procedure or Method (2017); Material Deformation (2976)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2024
Type of Reportable Event Serious Injury
Event or Problem Description
An attempt was made to use one resolute onyx coronary drug eluting stent (des) to treat a non-tortuous, non-calcified lesion located in the mid posterior tibial artery.The device was not inspected.Negative prep was performed with no issues.The lesion was pre-dilated.The device did not pass through a previously deployed stent.Resistance was not encountered when advancing the device.Excessive force was not used during delivery.It was reported that a balloon burst or leak occurred during stent deployment. nominal pressure was applied prior to the burst or leak, but the balloon would not inflate.Two inflations were applied to the device prior to the issue.It is believed that there was a pin size hole in the balloon of the stent.The patient is alive with no injury.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: the lesion had 10% stenosis.It was later confirmed, that a balloon burst did not occur.However, inflation issues occurred, on the first inflation on the balloon.And the first inflation never reached the minimum pressure. there was only partial inflation of the balloon.The device was not moved or repositioned, while inflated.The same inflation device was successfully used with other devices.The stent was implanted and needed to be post-dilated with another balloon.The patient is alive with no further injury.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank, because the information is currently, unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product analysis summary: the device returned to medtronic for evaluation.Kinks were evident to the hypotube.The device returned with balloon folds intact/expanded.The balloon failed negative prep.On pressurization of the device, liquid was observed exiting the balloon distal cone.The balloon failed to maintain pressure.Upon visual inspection of the device, a pin hole was observed on the balloon material, on the balloon distal cone.No other damage evident to the remainder of the device.Correction: b.Adverse event or product updated to adverse event, product problem.H.1 type of reportable event: serious injury ticked.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: annex d codes.Correction: annex a code.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESOLUTE ONYX RX
Common Device Name
STENT, CORONARY, DRUG-ELUTING
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key19210986
Report Number9612164-2024-02046
Device Sequence Number8193981
Product Code NIQ
UDI-Device Identifier00643169557024
UDI-Public00643169557024
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P160043
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/13/2026
Device Model NumberRONYX30030UX
Device Catalogue NumberRONYX30030UX
Device Lot Number0011636538
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/25/2024
Supplement Date Received by Manufacturer05/03/2024
05/16/2024
05/29/2024
Initial Report FDA Received Date04/30/2024
Supplement Report FDA Received Date05/08/2024
05/29/2024
05/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/14/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-