| Catalog Number |
8065771541 |
| Medical Device Problem Code |
Dull, Blunt (2407)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/16/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A healthcare professional reported that a female patient underwent cataract surgery on her right eye, during which the ophthalmic knife failed to make an incision.The surgery was completed on the same day using the same knife with increased force, and then switched to a new knife.The surgery proceeded without complications, and no medical or surgical intervention was necessary.However, the patient's outcome was currently unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
One opened knife, in a blister was received for the report of blunt knife.Sample was visually inspected and found to be nonconforming, blade was observed to have a bent tip.Penetration testing could not be performed due to the damage of the sample.A review of the device history record traceable to the reported lot numbers, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria the exact root cause could not be determined from the investigation performed.The damage to the returned sample is consistent with damage that can occur when the blade contacts a hard surface such as the protective blade tray when product is improperly removed or inserted after use, or improper handling, or contact with another instrument during surgery or set-up.The damage seen on the returned complaint sample could have contributed to the customers reported issue of blunt knife.The exact root cause for the damaged knife sample is unknown, therefore specific action with regards to this complaint cannot be taken.A nonconformance investigation is currently in progress to investigate the trend identified for dull cutting instruments.All knives are 100% inspected by trained operators using a minimum of 10x magnification during manufacturing.Any nonconformance, such as the damaged tip exhibited on the returned opened sample is removed from the lot and scrapped.Functional penetration testing is performed and monitored during the finishing process to ensure the sharpness of the product.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.The manufacturer internal reference number is: (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information is provided in d.4 the manufacturer internal reference number is:(b)(4).
|
| |
|