• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DAVOL INC., SUB. C.R. BARD, INC. -1213643 BARD SOFT MESH; SURGICAL MESH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DAVOL INC., SUB. C.R. BARD, INC. -1213643 BARD SOFT MESH; SURGICAL MESH Back to Search Results
Catalog Number 0117008
Medical Device Problem Code Patient-Device Incompatibility (2682)
Health Effect - Clinical Codes Seroma (2069); Hernia (2240); Insufficient Information (4580)
Date of Event 06/29/2021
Type of Reportable Event Serious Injury
Event or Problem Description
Attorney alleges multiple patients he represents have experience unspecified injury and allege they may have required additional medical/surgical treatment.Attorney alleges this patient was treated with a soft mesh.Case-specific details not provided.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges unspecified adverse patient outcome associated with the use of hernia mesh.The degree to which the device used to treat the patient may have caused or contributed to a post-op complication is unknown.No lot number has been provided, a review of the manufacturing records is not possible.Note, the date of event is considered to be a best estimate as 10-apr-2024 based on the date of awareness.Should additional information be provided, a supplemental emdr will be submitted.Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : not returned.
 
Additional Manufacturer Narrative
No conclusions can be made.The patient's attorney alleges unspecified adverse patient outcome associated with the use of hernia mesh.The degree to which the device used to treat the patient may have caused or contributed to a post-op complication is unknown.No lot number has been provided, a review of the manufacturing records is not possible.Addendum: h11: this supplemental emdr is submitted to document additional information provided and to correct the event date.Based on the additional information received, no conclusions can be made.Per medical records review, about a month post implant of bard soft mesh, patient was diagnosed with hernia recurrence and underwent a revision procedure.Post revision the patient developed a seroma.The instructions-for-use supplied with the device lists hernia recurrence and seroma as possible complications.Review of manufacturing records confirms product was manufactured to specification.To date, this is the only reported complaint for this manufacturing lot of (b)(4) units released for distribution in september 2019.Corrected fields: b2 (event attributed to), b3 (date of event).Note: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event or Problem Description
Attorney alleges multiple patients he represents have experience unspecified injury and allege they may have required additional medical/surgical treatment.Attorney alleges this patient was treated with a soft mesh.Case-specific details not provided.Addendum per information provided by patient's attorney: (b)(6) 2021 - patient was diagnosed with incarcerated indirect right inguinal hernia thereby underwent repair with the implant of bard soft mesh.(b)(6) 2021 - patient was diagnosed with recurrent incarcerated right inguinal hernia and pain thereby underwent open preperitoneal mesh repair.(note, no operative notes provided).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD SOFT MESH
Common Device Name
SURGICAL MESH
Manufacturer (Section D)
DAVOL INC., SUB. C.R. BARD, INC. -1213643
100 crossings blvd.
warwick RI 02886
Manufacturer (Section G)
BARD SHANNON LIMITED -3005636544
san geronimo industrial park
lot #1, road #3, km 79.7
humacao PR 00791
Manufacturer Contact
andrew topoulos
100 crossings blvd.
warwick, RI 02886
8005566756
MDR Report Key19211760
Report Number1213643-2024-00200
Device Sequence Number6960833
Product Code FTL
UDI-Device Identifier00801741030857
UDI-Public(01)00801741030857(17)240828(10)HUDV0669
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K052155
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source Other,Foreign,Consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/28/2024
Device Catalogue Number0117008
Device Lot NumberHUDV0669
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2024
Supplement Date Received by Manufacturer07/30/2024
Initial Report FDA Received Date04/30/2024
Supplement Report FDA Received Date08/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured09/14/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Other;
Patient Age68 YR
Patient SexMale
Patient Weight71 KG
-
-