Catalog Number 550036100
Medical Device Problem Code Device Dislodged or Dislocated (2923)
Health Effect - Clinical Codes Joint Dislocation (2374); Joint Laxity (4526)
Date of Event 04/17/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.1.Quantity of manufactured- (b)(4).2.Date of manufacturing- 2022-08-30 (per the label control form).3.Ant anomalies or deviations identified in dhr- i reviewed the manufacturing dhr¿s and did not identify any anomalies or deviations.There were some scrapped parts at the machine operation due to engraving not being deep enough and at the finishing operation due to material finish.These are not outside of normal manufacturing practice.All inspection documents have been reviewed and meet defined requirements.4.Expiry date- 2027-07-31.5.Ifu reference- no physical ifu referenced on packaging specification outside of what was on the label.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : 1.Quantity of manufactured- (b)(4).2.Date of manufacturing- 2022-08-30 (per the label control form).3.Ant anomalies or deviations identified in dhr- i reviewed the manufacturing dhr¿s and did not identify any anomalies or deviations.There were some scrapped parts at the machine operation due to engraving not being deep enough and at the finishing operation due to material finish.These are not outside of normal manufacturing practice.All inspection documents have been reviewed and meet defined requirements.4.Expiry date- 2027-07-31 5.Ifu reference- no physical ifu referenced on packaging specification outside of what was on the label.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
It was reported that patient was revised due to dislocation.Patient had original surgery done at hospital on (b)(6) 2024.Unfortunately they dislocated and needed to be revised.Surgeon removed the poly, shell, and glenosphere.A bigger glenosphere was implanted to increase stability.There was no adverse events or delays took place.Doi: (b)(6) 2024; dor: unknown; affected side: left shoulder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REVERSE LINER 36+0 R
Common Device Name
SHOULDER IMPLANT - HUMERAL CUP
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key19212964
Report Number1818910-2024-09445
Device Sequence Number20568263
Product Code PHX
UDI-Device Identifier10603295546788
UDI-Public10603295546788
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K212737
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number550036100
Device Lot NumberMI124724
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/16/2024
Supplement Date Received by Manufacturer05/09/2024
Initial Report FDA Received Date04/30/2024
Supplement Report FDA Received Date05/13/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured08/30/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GLENOSPHERE 36+8; REVERSE HUMERAL SHELL L 40+0
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
Patient SexUnknown
-
-