| Catalog Number |
ER8075 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/01/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The lot number was reviewed for complaint trend, nonconforming report and capa.Devices met specification prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed in veeva to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was removed and replaced due to a tubing leak.
|
| |
|
Additional Manufacturer Narrative
|
|
Reservoir was received for evaluation.No functional abnormalities were noted with the reservoir.The information received indicated the device had a tubing leakage, but because no functional abnormalities were noted with the returned component, the complaint could not be confirmed as reported.A review of the device history record confirmed the devices from this lot met all specifications prior to release.No trends were noted for complaints and there were no nonconforming reports or capas that were confirmed to be associated.
|
| |
|
Event or Problem Description
|
|
According to the available information, the device was removed and replaced due to a tubing leak.
|
| |
|
Search Alerts/Recalls
|