| Model Number |
93505 |
| Medical Device Problem Codes |
Break (1069); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
As reported, during use in patient, the manifold of this co-set broke.There is no allegation of patient injury.Patient demographics were not available.
|
| |
|
Additional Manufacturer Narrative
|
|
The product is expected to be returned for analysis; however, it has not yet been received.Upon the return of the product an investigation will be completed to consider any potential factors that may have contributed to this complaint and a supplemental report will be sent with the investigation results.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|
| |
|
Additional Manufacturer Narrative
|
|
Added information to section d9 and h6 (type of investigation).Updated section h6 (component code), h6 (investigation findings) and h6 (investigations conclusions).The device was received for a full evaluation.The report of co-set broken was confirmed.The male luer of coset flow-through housing had been completely broken.Broken male luer was not returned.Cross surface of broken luer was rough and uneven.No other visible damage was observed from returned unit.The manufacturing records were reviewed for the lot involved and there is no indication of a related nonconformance.All process parameters were met and inspections passed successfully.The cause of the issue could be traced to device desing.The male luer meets all specifications and aligns with customer requirements, however the issue is caused when the user puts more stress on the male luer than intentened.Corrective actions are being implemented in order to eliminate the cause of the non-conformity and prevent recurrence of this type of complaint.
|
| |
|
Additional Manufacturer Narrative
|
|
Updated section h6 (device code) to better reflect the nature of the event.
|
| |
|