| Catalog Number |
PXB35-09-27-080 |
| Medical Device Problem Code |
Burst Container or Vessel (1074)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/14/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Physician attempted to use a visipro balloon expandable stent during patient treatment.There were no abnormalities reported in relation to anatomy.There was no damage noted to packaging (i.E.Shelf carton, hoop/tray).A balloon burst/leak was reported during stent deployment.A replacement medtronic device was used to complete procedure.No patient injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis the device was returned in a pre-inflated state and with the stent on the balloon.The returned stent was confirmed to be 27mm a 20ml water filled syringe was used to flush the device and a steady stream of water was observed exiting the tip a 0.014¿ guidewire was loaded through the tip, and negative prep was performed with no bubbles present an indeflator with pressure gauge was used to inflate the balloon to its nominal pressure of 10atms and its rated burst pressure (rbp) 12atms without any issues and the balloon held pressure the deployed stent was confirmed as 27mm medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|