• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION ENDOVIVE SAFETY PEG KIT; TUBES, GASTROINTESTINAL (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION ENDOVIVE SAFETY PEG KIT; TUBES, GASTROINTESTINAL (AND ACCESSORIES) Back to Search Results
Model Number M00566470
Medical Device Problem Code Obstruction of Flow (2423)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Block h6: imdrf device code a1409 captures the reportable event of peg tube obstructed.Block h10: the customer reported that the device was disposed and will not be returned.As such, physical analysis has not been conducted in our laboratory.However, a device history review was performed and determined that, based on the nature of the reported event and the reported device lot number, the peg tube obstruction was likely the result of the manufacturing process.With all the available information, although the device has not been returned, boston scientific corporation (bsc) determined that the peg tube obstruction was likely caused by adhesive inside the hypo-tube of the transition joint.Adhesive is used during assembly, and it is likely that the obstruction was the result of incorrect placement or excessive use of adhesive, resulting in obstruction of the hypo tube.Bsc initiated a corrective and preventive action (capa) investigation to determine the root cause of this issue.Investigation identified a potential scenario in which the tip of the glue dispenser touches the threaded tip of the barb as it moves into position for glue and torque at the mini bolster.This could cause glue to migrate into the lumen of the barb component and cause a blockage.An initial containment was implemented on 05dec2023, and the pneumatic flow valve of the equipment was adjusted to ensure that the glue dispenser does not move into position until the barb is in the correct position, preventing the potential for this issue to reoccur.Additional corrections were implemented on 05feb2024 to add 100% pin gage inspection after the mini bolster step for push method device types.An investigation to address this problem has been completed.A labeling review was performed, and from the information available, this device was used per the instructions for use (ifu)/product label.
 
Event or Problem Description
Note: this report pertains to the third of three endovive safety peg kits push method that were used in the same patient and procedure.It was reported to boston scientific corporation that an endovive safety peg kit push method was attempted to be used during a percutaneous endoscopic gastrostomy (peg) tube placement procedure performed on (b)(6) 2024.During the procedure, the guidewire could not pass through the peg tube, the inner lumen was obstructed.They looked into the tube and could not see an opening.Seems like it was sealed at the junction point of the leading tapered tube and g-tube.The customer reported that two additional peg kits were opened and attempted to be used but the same problem occurred.The procedure was completed with another push method peg kit.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOVIVE SAFETY PEG KIT
Common Device Name
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
500 commander shea boulevard
quincy MA 02171
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key19214731
Report Number3005099803-2024-01962
Device Sequence Number15488520
Product Code KNT
UDI-Device Identifier08714729748397
UDI-Public08714729748397
Combination Product (Y/N)Y
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K031538
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Nurse
Remedial Action Recall
Type of Report Initial
Report Date (Section B) 04/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/31/2025
Device Model NumberM00566470
Device Catalogue Number6647
Device Lot Number0033112086
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/09/2024
Initial Report FDA Received Date04/30/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/02/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
97172867-FA
Patient Sequence Number1
Patient SexUnknown
-
-