| Catalog Number |
7209485 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/27/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.One of the suture retrievers had the snare fractured away from the shaft and hub.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the reported event include an application of unintended inappropriate or excessive force to the device, or an impact event inconsistent with normal use.No containment or corrective actions are recommended at this time.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case-(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that during a knee surgery, the lasso of the set meniscus mender was introduced into the needle put into the knee of the patient to get back a thread.By trying to remove the system, the tip of the lasso remained in to the patient knee.The surgeon succeeded to get it back (no damage for the patient).The procedure was completed with a delay of 15 minutes.There is no information available of the back up of the set meniscus mender.No further complications were reported.
|
| |
|
Search Alerts/Recalls
|