• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES SAPIEN XT TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES SAPIEN XT TRANSCATHETER HEART VALVE; AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number 9300TFX23A
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Dyspnea (1816); Insufficient Information (4580)
Date of Event 04/10/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Per the instructions for use (ifu), structural valve deterioration (wear, fracture, calcification, leaflet tear/tearing from the stent posts, leaflet retraction, suture line disruption of components of a prosthetic valve, thickening, stenosis), and device degeneration are known potential risks associated with bioprosthetic heart valves.The ifu cautions that accelerated deterioration of the valve may occur in patients with an altered calcium metabolism.Long-term durability has not been established for the valve.Regular medical follow-up is advised to evaluate valve performance.Structural valve deterioration (svd) may be manifested as stenosis with thickened leaflets.Svd may be mild and not require any intervention or it may be moderate to severe.It can cause the heart to work harder to eject blood from the ventricle.Depending on the severity it could be an indication for valve replacement or medical intervention.Tissue calcification is a very common failure mode.The mechanisms for bioprosthetic heart valve tissue calcification are not yet fully understood.Many factors can contribute to the onset and propagation of calcification including patient-related (e.G., patient age, disease state, immune status, and other co-morbidities), pharmacological, and intrinsic properties of the valve itself.It is widely understood that patients with chronic renal disease and prior history of calcific stenosis of the native valve may be predisposed to bioprosthetic calcification.During the manufacturing process, all sapien thv valves are 100% visually inspected for defects and 100% functionally tested for proper coaptation under physiological backpressure conditions prior to release for distribution.Therefore, it is highly unlikely that a manufacturing defect or device malfunction contributed to the event.In this case, there was no allegation or indication a product malfunction contributed to this adverse event.Investigation results suggest/indicate the cause of the event is unknown due to the limited information available.A review of edwards lifesciences risk management documentation was performed for this case.The reported event is an anticipated risk of the transcatheter heart valve procedure, additional assessment of this adverse event is not required at this time.Complaint histories for all reported events are reviewed against trending control limits monthly, and any excursions above the control limits are assessed and documented as part of this monthly review.As such, neither a product risk assessment, nor corrective or preventative actions are required at this time.
 
Event or Problem Description
As reported by a field clinical specialist (fcs), approximately 8 years and 7 months post-implant of a 23 mm sapien xt valve in the aortic position via transfemoral approach.The valve failed due to stenosis.The baseline peak 100 mmhg and mean 60 mmhg.The current peak/mean gradient is 10 mmhg cath gradient.The patient was symptomatic.The team do not believe the failure to be early as valve was implanted 9 years ago.A valve in valve was successfully performed.
 
Additional Manufacturer Narrative
A supplemental mdr is being submitted for correction and additional information.The following sections of this report have been updated: corrected h.6 clinical code.Per the field clinical specialist (fcs), the patient presented with dyspnea and was hospitalized for heart failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SAPIEN XT TRANSCATHETER HEART VALVE
Common Device Name
AORTIC VALVE, PROSTHESIS, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key19215701
Report Number2015691-2024-03240
Device Sequence Number8105427
Product Code NPT
UDI-Device Identifier00690103192070
UDI-Public(01)00690103192070(17)170604
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
P140031
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/20/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/04/2017
Device Model Number9300TFX23A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2024
Supplement Date Received by Manufacturer05/14/2024
Initial Report FDA Received Date04/30/2024
Supplement Report FDA Received Date05/20/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexFemale
-
-