• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD-SOLIS AMBULATORY INFUSION PUMP; PUMP, INFUSION, PCA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD-SOLIS AMBULATORY INFUSION PUMP; PUMP, INFUSION, PCA Back to Search Results
Model Number 2110
Device Problem Device Alarm System (1012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2024
Event Type  malfunction  
Manufacturer Narrative
B3: event date unknown.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional information becomes available.
 
Event Description
It was reported that the device exhibited error code 44510.The event was an out-of-box failure and occurred during testing.There was no patient involvement and no patient harm.
 
Manufacturer Narrative
One device was received for evaluation.Visual inspection found a damaged keypad, damaged lens, and a damaged pca connector.There was no applicable evidence in the event history log.Functional testing was able to duplicate the reported problem.It was determined that the faulty pwa board was the root cause and was replaced.The service history review had no indication that the complaint was related to a service of the device within the review period.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD-SOLIS AMBULATORY INFUSION PUMP
Type of Device
PUMP, INFUSION, PCA
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan ln n
minneapolis, MN 55442
2247062300
MDR Report Key19217180
MDR Text Key341434505
Report Number3012307300-2024-03195
Device Sequence Number1
Product Code MEA
UDI-Device Identifier10610586038778
UDI-Public10610586038778
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K130394
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/30/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2110
Device Catalogue Number21-2111-0300-01
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer04/23/2024
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 04/02/2024
Initial Date FDA Received04/30/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received05/30/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/22/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-