NELLCOR PURITAN BENNETT MEXICO SA NELLCOR; OXIMETER Back to Search Results
Model Number MAX-I-I
Medical Device Problem Codes Erratic or Intermittent Display (1182); Failure to Analyze Signal (1539); Low Readings (2460)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/02/2024
Type of Reportable Event Malfunction
Event or Problem Description
According to the reporter, during use, the product gave low oxygen saturation (spo2) readings, erratic spo2 readings, and no spo2 readings.The sensor site was checked/changed every 4-6 hours, wrap was used to hold the sensor in place, and led lighting was used. there was no patient injury.
 
Additional Manufacturer Narrative
D10 concomitant product: max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h); max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h); max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h); max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h); max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h); max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h); max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h); max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h); max-i-i max inf o2 sensor i20 3-20 kg (lot#: 223110045h).Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
H3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the device found no notable conditions.Functional testing found that the detector's internal white wire was broken close to the sensor.It was reported that during use, the product gave low spo2 readings, erratic spo2 readings, and no spo2 readings.The reported issues were confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NELLCOR
Common Device Name
OXIMETER
Manufacturer (Section D)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX  22225
Manufacturer (Section G)
NELLCOR PURITAN BENNETT MEXICO SA
boulevard insurgentes 19030
tijuana 22225
MX   22225
Manufacturer Contact
justin ellis
8200 coral sea st ne
mounds view, MN 55112
7635265677
MDR Report Key19217725
Report Number1282497-2024-00079
Device Sequence Number15503573
Product Code DQA
UDI-Device Identifier20884522041277
UDI-Public20884522041277
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K012891
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2025
Device Model NumberMAX-I-I
Device Catalogue NumberMAX-I-I
Device Lot Number223110045H
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/07/2024
Supplement Date Received by Manufacturer06/21/2024
Initial Report FDA Received Date04/30/2024
Supplement Report FDA Received Date07/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/07/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE NOTE ON H11
Patient SexMale
-
-