| Model Number |
VTICM5_13.2 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Codes |
Iritis (1940); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/09/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - work order search: no similar complaint was reported for units within the same lot.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm vticm5_13.2 implantable collamer lens, -10.50/+3.0/090 (sphere/cylinder/axis), into the patient`s right eye (od) on (b)(6) 2024.The patient had difficulty reading phone and computer; had constant dry eye, causing pain and irritation.Also had intermittent cloud-like film.The lens remained implanted.This was the replacement lens for mfr.Report # 2023826-2024-01675.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: post-op, the icl was not sitting properly due to patients anatomy and there was pain and swelling.Iritis was observed and patient was started on 1% prednisolone acetate q2h for 1 day then qid until next appointment.Exam noted "area of iris-icl touch sup" and "icl, vault-0.75:1, 1+ pigment".No cell or flare.At post-op visit on (b)(6) 2024, used 3 mirror gonio to examine icl.The icl appeared to have rotated sitting @ 165 degrees.Iritis had resolved.On (b)(6) 2024 the lens was explanted.At 1 day post-op, ucdva was cf@1ft and iop was 10.H6: health effect impact code - additional medications: 4644 - prednisolone acetate.Claim # (b)(4).
|
| |
|
Search Alerts/Recalls
|