| Model Number |
VICM5_13.2 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A2: unk.A3: unk.A4: unk.A5: unk.A6: unk.B3: unk.D6a: unk.D6b: unk.H6 - work order search: no similar complaint was reported for units within the same lot.Claim # (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated the surgeon implanted a 13.2mm vicm5_13.2 implantable collamer lens -11.50 diopter, into the patient`s left eye (os).The lens was explanted due to the surgeon noted a defect on the lens when it was implanted into the eye.The lens was replaced with another same model/length lens.Additional information has been requested but none has been forthcoming.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: device evaluation: the lens was returned dry, in a plastic bag, with residue on product.Visual inspection found the lens optic and haptics torn.There was residue throughout the lens.(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected data: b1: product problem b5: the reporter indicated the surgeon inserted and removed a 13.2mm vicm5_13.2 implantable collamer lens, -11.50 diopter, within the same surgery due to the surgeon noted a defect on the lens when it was implanted into the patient's left eye (os).The lens was replaced with another same model/length lens.Additional information has been requested but none has been forthcoming.H1: malfunction h6: adverse event problem: health effect - impact code (f): 2199.H6: adverse event problem: medical device problem code: (a): 3189.Claim # (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
H6: device history record (dhr) review: the dhr review indicated that the product has been manufactured within the established process parameters and that there is no indication that the manufacturing and/or processing of the device contributed to the complaint issue.(b)(4).
|
| |
|
Search Alerts/Recalls
|