MAUDE Adverse Event Report: EXACTECH, INC. EQUINOXE CAGE GLENOID M, POST AUG, RIGHT; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. EQUINOXE CAGE GLENOID M, POST AUG, RIGHT; PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED Back to Search Results
Catalog Number 314-13-33
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Loosening of Implant Not Related to Bone-Ingrowth (4002)
Health Effect - Clinical Codes Failure of Implant (1924); Insufficient Information (4580)
Date of Event 02/28/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(d10) concomitant device(s): 300-62-01 - stemless humeral comp integrip, cage, size 1 6348741, 310-62-50 - stemless humeral head 50mm x 16mm x beta 6160584, (h3) pending evaluation.
 
Event or Problem Description
As reported by the equinoxe shoulder study, the 69-year-old white female had a right tsa on (b)(6) 2020.The patient presented with aseptic glenoid loosening on (b)(6) 2024.The patient underwent a revision surgery on (b)(6) 2024 and the outcome of this event is considered resolved.The case report form indicates that this event is possibly related to the device and definitely not related to the procedure.This event report was received through clinical data collection activities and no device return is anticipated.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.The reason for the revision reported cannot be confirmed from the information provided but may be the result of an insufficient bond between the glenoid component and the bone, which led to aseptic (non-infected) glenoid loosening.Potential contributions of user-related considerations to the event could not be assessed as the devices were not available for evaluation and images, radiographs, and relevant clinical information were not provided.If any further information is obtained that would change or alter any information provided, a supplemental report will be filed accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUINOXE CAGE GLENOID M, POST AUG, RIGHT
Common Device Name
PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH
2320 nw 66th ct.
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th ct.
gainesville, FL 32653
3523771140
MDR Report Key19218038
Report Number1038671-2024-01018
Device Sequence Number6904224
Product Code KWS
UDI-Device Identifier10885862201010
UDI-Public10885862201010
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K113309
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2024
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 02/13/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date11/04/2024
Device Catalogue Number314-13-33
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/12/2024
Supplement Date Received by Manufacturer02/11/2025
Initial Report FDA Received Date04/30/2024
Supplement Report FDA Received Date02/13/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured11/08/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1415-2024
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H10.
Patient Age69 YR
Patient SexFemale
Patient Weight126 KG
Patient RaceWhite
-
-