• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) CR2 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK(R) CR2 DEFIBRILLATOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number CR2
Medical Device Problem Code Inappropriate/Inadequate Shock/Stimulation (1574)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/09/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Stryker contacted the customer to request additional information on the patient.The customer provided stryker with the available patient information.Patient fields in which information is not provided were intentionally left blank.The customer will receive a replacement device.Stryker continues to investigate the reported issue and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
 
Event or Problem Description
The customer contacted stryker to report that their device delivered autonomously shock when users were performing cpr.In this state, wrong defibrillation therapy may be delivered.There were no adverse consequences to the patient as a result of the reported event.
 
Event or Problem Description
The customer contacted stryker to report that their device automatically delivered shock to the patient even though the patient was awake with stabile vital signs.There were no adverse consequences to the patient as a result of the reported event.
 
Additional Manufacturer Narrative
Executive summary of initial mdr indicates: the customer contacted stryker to report that their device delivered autonomously shock when users were performing cpr.In this state, wrong defibrillation therapy may be delivered.There were no adverse consequences to the patient as a result of the reported event.Executive summary of initial mdr should indicate: the customer contacted stryker to report that their device automatically delivered shock to the patient even though the patient was awake with stabile vital signs.There were no adverse consequences to the patient as a result of the reported event.
 
Additional Manufacturer Narrative
The product assessment center (pac) technician evaluated the customer's device and was not able to verify the reported issue.A root cause of the reported issue could not be determined.The device was archived by stryker.Section a4 weight units of initial mdr indicates: kilogram(s).Section a4 weight units of initial mdr should indicate: pounds.
 
Event or Problem Description
The customer contacted stryker to report that their device automatically delivered shock to the patient even though the patient was awake with stabile vital signs.There were no adverse consequences to the patient as a result of the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK(R) CR2 DEFIBRILLATOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer (Section G)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
Manufacturer Contact
brian blakeslee
11811 willows road ne
redmond, WA 98052
4258674000
MDR Report Key19218279
Report Number0003015876-2024-01385
Device Sequence Number20575276
Product Code MKJ
UDI-Device Identifier00883873824115
UDI-Public00883873824115
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P170018
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberCR2
Device Catalogue Number99512-001263
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/10/2024
Supplement Date Received by Manufacturer05/16/2024
05/21/2024
Initial Report FDA Received Date04/30/2024
Supplement Report FDA Received Date05/16/2024
05/23/2024
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured02/02/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient Age76 YR
Patient SexMale
Patient Weight78 KG
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-