| Model Number |
VTICM5_13.2 |
| Medical Device Problem Codes |
Adverse Event Without Identified Device or Use Problem (2993); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/07/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
H6 - type of investigation - lens work order search: no similar complaint type events reported for units within the same lot.Claim# (b)(4).
|
| |
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm vticm5_13.2 implantable collamer lens of - 10.50/1.5/091 (sphere/cylinder/axis) was implanted into the patient's left eye (os) on (b)(6) 2024.The reporter reports observation of low vault without rotation.Lens remains implanted.If additional information is received a supplemental medwatch report will be submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
B5: per updated information the lens was exchanged with a longer length lens and the problem resolved.Manufacturer narrative - secondary surgical intervention to remove/replace/reposition the lens is identified in the labeling as a known potential adverse event following icl implantation.Claim # (b)(4).
|
| |
|