WARSAW ORTHOPEDICS ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM; APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD Back to Search Results
Model Number 7200055
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/04/2024
Type of Reportable Event Malfunction
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient having anterior cervical discectomy and fusion (acdf).Levels implanted: c4-7.Pre-operative diagnosis for this procedure was myelopathy.It was reported that the device locking mechanism came off.There were no patient symptoms reported.There were no further complications reported regarding the event.Additional information was received from the manufacturer representative that the locking mechanism broke off.
 
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Product analysis of part # 7200055 ; lot # unknown visual inspection revealed one of the locking mechanisms is missing.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATLANTIS ANTERIOR CERVICAL PLATE SYSTEM
Common Device Name
APPLIANCE, FIXATION, SPINAL INTERVERTEBRAL BOD
Manufacturer (Section D)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer (Section G)
WARSAW ORTHOPEDICS
2500 silveus crossing
warsaw IN 46582
Manufacturer Contact
glen belmer
1800 pyramid place
memphis, TN 38132
6122713209
MDR Report Key19219358
Report Number1030489-2024-00367
Device Sequence Number8194229
Product Code KWQ
UDI-Device Identifier00613994625304
UDI-Public00613994625304
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K021461
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 05/31/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7200055
Device Catalogue Number7200055
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2024
Supplement Date Received by Manufacturer05/08/2024
Initial Report FDA Received Date05/01/2024
Supplement Report FDA Received Date05/31/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age51 YR
Patient SexFemale
-
-