MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HUNTINGTON CLAREON VIVITY TORIC EXTENDED VISION UV ABSORBING IOL; LENS, INTRAOCULAR, TORIC OPTICS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HUNTINGTON CLAREON VIVITY TORIC EXTENDED VISION UV ABSORBING IOL; LENS, INTRAOCULAR, TORIC OPTICS Back to Search Results
Model Number CCWET3
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Disturbances (2140)
Date of Event 12/14/2023
Type of Reportable Event Serious Injury
Event or Problem Description
A facility representative reported that following an intraocular lens (iol) implant procedure, the patient experienced glare.The lens was exchanged for another company lens model 11 months following the initial procedure.Clinical reason for explant was mentioned as mechanical complication of iol.Additional information has been received that there was no patient harm reported.There are two medical device reports associated with this file.This report is associated with the right eye.
 
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Additional Manufacturer Narrative
The product was not returned.Product history records were reviewed and the documentation indicated, the product met release criteria.The account indicated, the use of a qualified cartridge, handpiece and viscoelastic.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100% assessment of the power and optical resolution, during the manufacturing process in order to determine acceptability, per the lens model and diopter.Information was provided that the toric, company (1.50cyl), 20.0 diopter, (1.5 extended depth of focus) lens was replaced with a non-company, non-toric, 19.0 diopter, monofocal lens model.The account indicated, the product was not available to evaluate.Due to the exchange of a toric-extended-depth-of-field lens to a non-toric monofocal lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened, if new information is received.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLAREON VIVITY TORIC EXTENDED VISION UV ABSORBING IOL
Common Device Name
LENS, INTRAOCULAR, TORIC OPTICS
Manufacturer (Section D)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer (Section G)
ALCON RESEARCH, LLC - HUNTINGTON
6065 kyle lane
huntington WV 25702
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key19219594
Report Number1119421-2024-00823
Device Sequence Number8194730
Product Code MJP
UDI-Device Identifier00380652455282
UDI-Public00380652455282
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P190018
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Reporter Type Manufacturer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/09/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/06/2026
Device Model NumberCCWET3
Device Catalogue NumberCCWET3.200
Device Lot Number15351735
Was Device Available for Evaluation? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 04/03/2024
Supplement Date Received by Manufacturer06/10/2024
Initial Report FDA Received Date05/01/2024
Supplement Report FDA Received Date07/09/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured02/08/2022
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DUOVISC; MONARCH II LOADING FORCEPS; MONARCH III CARTRIDGE D; MONARCH III INJECTOR; ORA SYSTEM WITH VERIFEYE MESSAGING; ORA SYSTEM WITH VERIFEYE MESSAGING, CART
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-