ARTHREX, INC. UNIVERS II HUMERAL STEM 9MM; SHOULDR PROSTH HEMI- HUM UNCEM Back to Search Results
Model Number UNIVERS II HUMERAL STEM 9MM
Medical Device Problem Code Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on may 5, 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.Complaint not confirmed, the device fit with the humeral head returned and the taper lock was found to meet specifications.
 
Event or Problem Description
On (b)(6) 2021, it was reported by a sales representative via email that an ar-9100-09p humeral stem, an ar-9152-20p humeral head, and an ar-9100-07s univers apex humeral stem all failed to fit.This was discovered during a procedure on (b)(6) 2021.Surgeon completed case by using a larger stem.Additional information received 12/3/2021: this occurred during a total shoulder arthroplasty.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNIVERS II HUMERAL STEM 9MM
Common Device Name
SHOULDR PROSTH HEMI- HUM UNCEM
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19219776
Report Number1220246-2024-02657
Device Sequence Number18192308
Product Code KWS
UDI-Device Identifier00888867057104
UDI-Public00888867057104
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K071032
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2025
Device Model NumberUNIVERS II HUMERAL STEM 9MM
Device Catalogue NumberAR-9100-09P
Device Lot Number20.03423
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/01/2021
Initial Report FDA Received Date05/01/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/01/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-