|
Mdr 1219913-2024-00299 was initially filed on 01-may-2024.Siemens has concluded the problem investigation.A united states (us) customer reported observation of an elevated advia centaur progesterone (prge) result which was discordant relative to repeat testing.Prge quality control (qc) results were within expected ranges.Service inspection revealed an issue with a fluid leak affecting cuvettes at the wash station.Multiple fluidics components (pumps, tubing, manifold) were replaced as part of a routine service response, and the leak was resolved.Following the service intervention, prge performance has been acceptable, and no further issues were reported.The instrument is performing according to specifications.The customer is operational, and the reported issue was resolved by service part replacement.In section h6, the codes for investigation findings, and investigation conclusions were updated.
|