SMITHS MEDICAL ASD, INC. PNEUPAC PARAPAC PLUS 300 VENTILATOR; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Catalog Number P300NJJP
Medical Device Problem Code Failure to Cycle (1142)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 08/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
B3.Date of event: month and year of event have been provided, but day is unknown.H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Event or Problem Description
It was reported that in august 2020, the ventilator had continuous cycling that occurred during inspection every morning.Resolved by repeating on/off.In september 2020, it occurred immediately after patient use.Switch to bvm.There was patient involvement and no patient harm/adverse event reported.
 
Additional Manufacturer Narrative
D9: date returned to mfg.: 4/18/2024.H3 and h6.Evaluation codes: updated.One device was received.No abnormality related to the reported event was confirmed on the product's appearance.Functional testing was unable to confirm/duplicate the complaint.It is presumed that due to o-ring material/hardness of input manifold, inspiration and exhalation did not switch, and oxygen gas continued to come out of the outlet of main unit.The input manifold assembly was replaced as a preventive measure.The device passed all functional tests after repair.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC PARAPAC PLUS 300 VENTILATOR
Common Device Name
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan ln n
minneapolis MN 55442
Manufacturer Contact
reed covert
6000 nathan lane n
minneapolis, MN 55442
2247062300
MDR Report Key19220801
Report Number3012307300-2024-03215
Device Sequence Number17868465
Product Code BTL
UDI-Device Identifier35019315107270
UDI-Public(01)35019315107270(11)181221
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K123957
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 06/12/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberP300NJJP
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/04/2024
Supplement Date Received by Manufacturer05/13/2024
Initial Report FDA Received Date05/01/2024
Supplement Report FDA Received Date06/12/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/21/2018
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1275-2024
Patient Sequence Number1
Patient SexUnknown
-
-