| Catalog Number |
P300NJJP |
| Medical Device Problem Code |
Failure to Cycle (1142)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
08/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
B3.Date of event: month and year of event have been provided, but day is unknown.H3.Reason device not evaluated by mfg: other; device was not returned to manufacturer.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported that in august 2020, the ventilator had continuous cycling that occurred during inspection every morning.Resolved by repeating on/off.In september 2020, it occurred immediately after patient use.Switch to bvm.There was patient involvement and no patient harm/adverse event reported.
|
| |
|
Additional Manufacturer Narrative
|
|
D9: date returned to mfg.: 4/18/2024.H3 and h6.Evaluation codes: updated.One device was received.No abnormality related to the reported event was confirmed on the product's appearance.Functional testing was unable to confirm/duplicate the complaint.It is presumed that due to o-ring material/hardness of input manifold, inspiration and exhalation did not switch, and oxygen gas continued to come out of the outlet of main unit.The input manifold assembly was replaced as a preventive measure.The device passed all functional tests after repair.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.
|
| |
|