|
Mdr 1219913-2024-00300 was initially filed on 01-may-2024.Additional information became available after the initial filing, and siemens has concluded the problem investigation.Service inspection revealed an issue with a fluid leak affecting cuvettes at the wash station.Multiple fluidics components (pumps, tubing, manifold) were replaced as part of a routine service response, and the leak was resolved.Following the service intervention, thcg performance has been acceptable, and no further issues were reported.The instrument is performing according to specifications.No further evaluation of the device is required.In section h6, the codes for investigation findings, and investigation conclusions were updated.
|