| Catalog Number |
A041602510212 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Device Embedded In Tissue or Plaque (3165); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/24/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As a result of this malfunction, the potential for surgical intervention exists to preclude permanent damage to a body structure or permanent impairment of a body function as evidenced by previous reported events with similar files.This event, therefore, is reportable per 21cfr part 803.The device was not returned for evaluation.However, the lot number was provided and retained-product testing and/or dhr review are planned.The results will be submitted as they become available.
|
| |
|
Event or Problem Description
|
|
In this event it is reported that a protaper h-u 25mm/s2 *not ce* broke during use.The broken part was not retrieved.No injury occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information received: the broken part that had remained in the canal has been retrieved.This is a follow up report for this additional information.Correcting health effect clinical code from 3165 to 4582.This is a follow up report for this corrected information.
|
| |
|
Additional Manufacturer Narrative
|
|
Summary: involved product that broke during use was not returned and cannot be analyzed.Moreover, no unused file is available for evaluation.Nothing unusual to report was found during dhr review (batch #1746842).Root causes are not identified.We will track this kind of event and monitor the trend.
|
| |
|