MAUDE Adverse Event Report: ARTHREX, INC. FLIPCUTTER II, SHORT 9.5MM; MANUAL INSTR, GENERAL SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. FLIPCUTTER II, SHORT 9.5MM; MANUAL INSTR, GENERAL SURGICAL Back to Search Results
Model Number FLIPCUTTER II, SHORT 9.5MM
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/05/2024
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The contribution of the device to the reported event could not be determined as the device was not returned for evaluation.The root cause of the event could not be determined from the information available and without device evaluation.If the device becomes available for evaluation, a follow-up report will be submitted.
 
Event or Problem Description
It was reported that during a knee arthroscopy the tip of both flipcutters broke during drilling the femoral tunnel.There was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the different part number.It was not necessary to switch the surgical technique or do a second surgery.***update swit 19-apr-2024: the broken pieces were retrieved from patient.
 
Additional Manufacturer Narrative
The complaint allegation is confirmed.One unpackaged ar-1204as-95 batch number 2136121965 was received for evaluation.A visual inspection revealed that the cutter tip blades were broken off.Additionally, the outer tube's outside diameter is bent at the proximal end.Since the device is damaged, performing a functional test isn't possible.A use error is the most likely cause for the reported and observed failure.Mechanical damage to the device, such as hitting the device with another device or object, prying/leveraging, or excessive bending forces applied during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLIPCUTTER II, SHORT 9.5MM
Common Device Name
MANUAL INSTR, GENERAL SURGICAL
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19222960
Report Number1220246-2024-02688
Device Sequence Number6941935
Product Code HWE
UDI-Device Identifier00888867004443
UDI-Public00888867004443
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial,Followup
Report Date (Section B) 09/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFLIPCUTTER II, SHORT 9.5MM
Device Catalogue NumberAR-1204AS-95
Device Lot Number2136121965
Was Device Available for Evaluation? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/08/2024
Supplement Date Received by Manufacturer08/22/2024
Initial Report FDA Received Date05/01/2024
Supplement Report FDA Received Date09/11/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured05/01/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-