Catalog Number 3507640MC
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Seroma (2069); Post Operative Wound Infection (2446)
Date of Event 04/03/2024
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: wound infection, seroma.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient implanted with a 640cc mentor memorygel breast implant experienced a wound infection and a seroma on the left side post-operatively.As a result, the patient underwent explantation on (b)(6) 2024.
 
Additional Manufacturer Narrative
Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation no apparent damage or visual anomalies were observed on the sm mod classic gel, 640cc returned device.Infection, manifested by swelling, tenderness, pain and fever, may appear in the immediate postoperative period or at any time after insertion of the implant.In the absence of classic symptoms, subacute or chronic infections may be difficult to diagnose.If infection does not subside promptly with the appropriate treatment, removal of the implant is indicated.Infection is a known complication associated with any surgery.Per a database query, this is the only reported complaint of infection against this lot number.The mentor microbiology department has provided information on the sterility lot for the implanted device indicating that it met all sterilization parameters required to provide a 10e-6 sterility assurance level prior to release for distribution.Product evaluation concluded the reported infection occurred from a source other than the device.Infection is a known complication associated with any surgery and is referenced in our current product insert data sheet.A seroma is a build-up of fluid around the implant.Symptoms from a seroma may include swelling, pain, and bruising.A seroma may occur soon after surgery, or any time later on, which would be referred to as a late seroma.While the body may absorb seromas, some may require surgery for drainage.You should consider the possibility of bia-alcl when a patient presents with late onset, persistent peri-implant seroma.In some cases, patients presented with capsular contracture or masses adjacent to the breast implant.When testing for bia-alcl, collect fresh seroma fluid and representative portions of the capsule, and sent to a laboratory with appropriate expertise for pathology test to rule out alcl, including immunohistochemistry testing for cd30 and alk (anaplastic lymphoma kinase).If your patient is diagnosed with peri-implant bia-alcl, develop an individualized treatment plan in coordination with a multidisciplinary care team.The national comprehensive cancer network (nccn) recommends surgical treatment that includes implant removal and complete capsulectomy ipsilaterally as well as contralaterally, where applicable.Report all confirmed cases of bia-alcl to the fda (https://www.Fda.Gov/safety/medwatch).Fda also recommends reporting cases of bia-alcl to the profile registry (https://www.Thepsf.Org/research/clinical-impact/profile.Htm) where you can submit more comprehensive data.Lastly, please notify mentor, because as the device manufacturer, we are collecting data on these cases as well.The american society of plastic surgeons recommends and encourages member surgeons to always submit breast implants, capsule, and effusion to pathology for examination.As part of mentor¿s quality process, all devices are manufactured, inspected, and released to approved specifications.Since no malfunction was observed during the investigation, no corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
kate karberg
3041 skyway circle north
irving, TX 75038
3035526892
MDR Report Key19225938
Report Number1645337-2024-05239
Device Sequence Number8305560
Product Code FTR
UDI-Device Identifier00081317004271
UDI-Public00081317004271
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number3507640MC
Device Lot Number9981875
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/11/2024
Supplement Date Received by Manufacturer05/12/2024
Initial Report FDA Received Date05/01/2024
Supplement Report FDA Received Date06/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/28/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age50 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-