| Catalog Number |
07K76-25 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/13/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The customer observed a falsely elevated architect prolactin result generated for a patient sample.The following data was provided (customer¿s reference range 1.2 - 29.9 ng/ml).(b)(6) 2024 sample id 10507185 initial result = 149.22 ng/ml, same sample repeated on (b)(6) 2024 and result = <0.6 ng/ml.No impact to patient management was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
An evaluation is in process.A final report will be submitted when the evaluation is complete.All available patient information was included.Additional patient details are not available.
|
| |
|
Additional Manufacturer Narrative
|
|
The complaint investigation included a review of data and information provided by the customer, search for similar complaints, ticket trending review, labeling review, device history record review, and in house testing of a retained reagent kit.Return testing was not completed as returns were not available.The ticket search by lot indicates that the complaint lot performs as expected for this product.Ticket trending review did not identify any trends.Device history review did not identify any non-conformances or deviations with the complaint lot and complaint issue.In-house accuracy testing was completed using panels which mimic patient samples using an in-house retained kit of the complaint lot.All specifications were met indicating that the lot is performing acceptably.Labeling was reviewed and sufficiently addresses the customer¿s issue.Based on the investigation, no systemic issue or deficiency of the architect prolactin reagent lot 59008ud00 was identified.
|
| |
|
Event or Problem Description
|
|
The customer observed a falsely elevated architect prolactin result generated for a patient sample.The following data was provided (customer¿s reference range 1.2 - 29.9 ng/ml) (b)(6) 2024 sample id (b)(6) initial result = 149.22 ng/ml, same sample repeated on (b)(6) 2024 and result = <0.6 ng/ml.No impact to patient management was reported.
|
| |
|