• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. G7 ACETABULAR LINER HIGH WALL 32MM E; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. G7 ACETABULAR LINER HIGH WALL 32MM E; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Pain (1994); Osteopenia/ Osteoporosis (2651)
Date of Event 04/11/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that approximately sixteen months post implantation, the patient was revised due to pain.There was no reported failure of any implants, but the surgeon noted possible malposition of the shell as contributing.The shell, liner, and head were revised.Attempts have been made and no further information has been provided.
 
Additional Manufacturer Narrative
(b)(4).D10: cat#: 51-113150, lot#: 6086695, tprlc 133 type1 bm so 15.0.G2: foreign ¿ australia.The customer has indicated that the product will not be returned to zimmer biomet for investigation as its location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Additional Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2024-01127, 0001825034-2024-01129.This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: b4; b5; g3; h2; h3; h6.No product was returned or pictures provided; visual and dimensional evaluations couldn¿t not be performed.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: the left hip arthroplasty is anatomically aligned.There is no fracture or dislocation.Bone quality appears osteopenic.The complaint cannot be confirmed.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No additional event information to report at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 ACETABULAR LINER HIGH WALL 32MM E
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key19228850
Report Number0001825034-2024-01128
Device Sequence Number5855732
Product Code PBI
UDI-Device Identifier00880304527034
UDI-Public(01)00880304527034(17)260212(10)6968586
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date02/12/2026
Device Model NumberN/A
Device Catalogue Number010000927
Device Lot Number6968586
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/11/2024
Supplement Date Received by Manufacturer07/16/2024
Initial Report FDA Received Date05/02/2024
Supplement Report FDA Received Date07/17/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured02/12/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention; Hospitalization;
Patient SexMale
-
-