| Model Number |
CCWET0 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Blurred Vision (2137)
|
| Date of Event |
03/04/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
A facility representative reported following an intraocular lens (iol) implant procedure, the patient experienced hazy distance vision.The lens was explanted and replaced with another lens in a secondary procedure.The clinical reason for the explant was mechanical defect.Additional information has been received and stated there was no patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is:(b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information was provided in h.3., h.6.And h.11.The product was not returned.Product history records were reviewed and the documentation indicated the product met release criteria.The account indicated the use of qualified associated products.The product investigation could not identify a root cause for the reported complaint.Each lens is subjected to a 100% assessment of the power and optical resolution during the manufacturing process in order to determine acceptability per the lens model and diopter.Information was provided that the toric, company model, 18.0 diopter, (1.5 extended depth of focus) lens was replaced with a non-company, 18.0 diopter, monofocal lens model.The product has not been received to evaluate.Due to the exchange of a extended-depth-of-field lens to a monofocal lens, this may suggest that the replacement lens model was an improved choice for the patient's vision needs.Due diligence has been performed in an attempt to obtain further information on this event.File will be reopened if new information or the sample is received.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|