| Model Number |
CCA0T0 |
| Medical Device Problem Codes |
Failure to Fold (1255); Failure to Eject (4010)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/04/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A non-health care professional reported that during an intraocular lens (iol) implant procedure, failed ejection.Lens did not folded properly.No patient harm was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was not returned for analysis.It is stated in the file that "a small amount of viscoelastic was used to prime the injector".It is unknown if the sufficient amount of viscoelastic was injected into the device.The root cause for the reported complaint could not be determined as no sample was returned for analysis.It is stated in the file that "a small amount of viscoelastic was used to prime the injector".It is unknown if the sufficient amount of viscoelastic was injected into the device.The ifu instructs: fill the device until ovd can be observed flowing to the nozzle tip.This will require approximately 0.28 ml of viscoelastic.If inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|