Model Number MR8-AA09
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Event or Problem Description
Repair request initiated for device with the report of overheating.No patient impact reported.Repair is being escalated to product event due to reason for the return.
 
Additional Manufacturer Narrative
H3: product analysis: the device was tested, and the maximum temperature was measured to be 147°f.The likely cause was identified as the bearings are corroded.It was also noted that the laser markings are faded, the input and output drive shafts are worn and corroded.Notified date: (b)(6) 2024.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MR8 ANGLED ATTACHMENTS
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer (Section G)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
Manufacturer Contact
glen belmer
7000 central avenue ne
ft. worth, TX 76137-3291
6122713209
MDR Report Key19230492
Report Number1625507-2024-00141
Device Sequence Number6834262
Product Code HBB
UDI-Device Identifier00643169912328
UDI-Public00643169912328
Combination Product (Y/N)N
Initial Reporter CountrySP
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 05/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMR8-AA09
Device Catalogue NumberMR8-AA09
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2024
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/09/2024
Initial Report FDA Received Date05/02/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/08/2021
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-