It was reported that during a tka surgery, the modular universal tibial impactor got fractured, after being used.The device fractured but remained in one piece so nothing remained in the patient.The procedure was resumed, without any delay, using the same device.Patient was not harmed as consequence of this problem.
|
|
H10: after further assessment of the information gathered by the manufacturer, it was identified that this event should be re-evaluated for mdr reporting.The original information stated that the modular universial tibial impactor got fractured.However, after further clarification and information received, it was determined that the issue does not meet the criteria to be reportable, since the device cracked but remained in one piece.Moreover, if the device malfunction were to recur, it would not be likely to cause or contribute to a death or serious injury.H3, h6: the device was not returned for evaluation.However, the photograph was reviewed, and revealed that the device cracked.The device shows signs of use.Device batch number was not provided, thus, an evaluation of the manufacturing records could not be performed.A review of complaint history of the previous 12 months revealed similar events for the listed device, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.Corrected data: h6 (medical device problem code).
|