• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE Back to Search Results
Model Number PCF-Q180AL
Medical Device Problem Code Failure to Clean Adequately (4048)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/17/2024
Type of Reportable Event Malfunction
Event or Problem Description
It was observed that during the device inspection, the colonovideoscope air channel was unable to work due to a clogged nozzle.There were no reports of patient involvement.
 
Additional Manufacturer Narrative
The device evaluation is ongoing.Should additional relevant information become available, a supplemental report will be submitted.
 
Additional Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 16 years since the subject device was manufactured.Based on the results of the investigation, a definitive root cause could not be established.The suggested phenomenon was confirmed, however, the foreign material in the nozzle was unable to be further specified.Moreover, no physical damage of the device was confirmed where the foreign material had remained, nor was any obvious deviation of reprocessing.Therefore, the cause of the remaining foreign material could not be presumed from the obtained information.The event can be detected and prevented in accordance with the following instructions for use (ifu): ifu states that detection method in evis exera ii gif/cf/pcf type 180 series operation manual chapter 3 preparation and inspection; ifu states that preventive measure in evis exera ii gif/cf/pcf type 180 series reprocessing manual chapter 5 reprocessing the endoscope (and related reprocessing accessories).It is recommended that the user facility contact olympus for repair when an abnormality of the device such as leakage/damage is detected ¿ the user facility is furthermore encouraged to contact olympus should they have questions or uncertain steps, so that observation/training may be provided.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19235021
Report Number9610595-2024-09107
Device Sequence Number12507581
Product Code FDF
UDI-Device Identifier04953170307720
UDI-Public04953170307720
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 06/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberPCF-Q180AL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/22/2024
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2024
Supplement Date Received by Manufacturer05/31/2024
Initial Report FDA Received Date05/02/2024
Supplement Report FDA Received Date06/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/13/2007
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-