| Model Number |
305U227 |
| Medical Device Problem Codes |
Calcified (1077); Gradient Increase (1270); Device Stenosis (4066); Intravalvular regurgitation (4068)
|
| Health Effect - Clinical Codes |
Aortic Valve Stenosis (1717); Aortic Valve Insufficiency/ Regurgitation (4450)
|
| Date of Event |
03/29/2024
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Event or Problem Description
|
|
Medtronic received information 13 years and 3 months post implant of this aortic bioprosthetic valve that the patient is being evaluated for a transcatheter valve-in-valve replacement.The reason for evaluation was reported as an increased peak gradient of 60mmhg and mean gradient of 36.2 mmhg, moderate to severe eccentric aortic regurgitation (ar), aortic stenosis, calcification, restricted motion, an aortic valve area of 1.3cm2, peak velocity of 3.87 m/s, and stroke volume index of 62.9.No intervention or adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Correction: b3- event date included medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
| |
|
Search Alerts/Recalls
|