AIZU OLYMPUS CO., LTD. EVIS GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-2T100
Medical Device Problem Code Device Reprocessing Problem (1091)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/29/2023
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device will not be returned to olympus for evaluation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, as the device was not returned, the root cause of the reported event is unable to be determined.However, it is presumable that the reported event is due to the user not following the instructions for use (ifu) which caused them to use the expired water filter passed the replacement period.The suggested event can be prevented by handling the device in accordance with the ifu: "chapter 7 routine maintenance 7.2 replacing the water filter (maj-824) replace the water filter at least once a month to prevent contamination of the rinse water.The water filter should also be replaced whenever an error code indicating water supply insufficiency [e01] is displayed.The water filter should be replaced by following the flow shown below." olympus will continue to monitor field performance for this device.
 
Event or Problem Description
It was reported that the subject device had been reprocessed with an endoscope reprocessor in which the water filter replacement deadline had expired by two years.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS GASTROINTESTINAL VIDEOSCOPE
Common Device Name
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key19237118
Report Number9610595-2024-09117
Device Sequence Number18148007
Product Code FDS
UDI-Device Identifier00724995091583
UDI-Public00724995091583
Combination Product (Y/N)N
Initial Reporter CountryEC
PMA/510(K) Number
K954451
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberGIF-2T100
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 01/04/2024
Initial Report FDA Received Date05/03/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured01/23/2019
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-