• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EXACTECH, INC. SPECIFIC DEVICE NOT REPORTED; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EXACTECH, INC. SPECIFIC DEVICE NOT REPORTED; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 02-012-35-2013
Medical Device Problem Codes Unstable (1667); Naturally Worn (2988)
Health Effect - Clinical Codes Failure of Implant (1924); Pain (1994); Joint Laxity (4526); Swelling/ Edema (4577)
Date of Event 09/07/2023
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The device was not returned for evaluation and no photographs or x-rays were provided for review; therefore the reported event cannot be confirmed through analysis.No operative notes were provided; therefore, a review of device usage and technique could not be performed.No information concerning patient conditions, co-morbidities, or other health or anatomical concerns was provided and it is therefore unknown if patient-related factors may have caused or contributed to the reported event.The cause of the reported event cannot be determined based on the information made available.Should additional, relevant information become available, a supplemental report will be filed accordingly.
 
Event or Problem Description
It was reported via legal documentation that approximately 64 months after a right total knee replacement procedure, the patient underwent a revision procedure to address polyethylene wear, pain, swelling, and instability.The revising surgeon noted bone damage and cancellous bone loss in the femoral condyle and tibial plateau.No further issues or complications were reported.No additional information is available.
 
Additional Manufacturer Narrative
D4: corrected the following: catalog number, expiration date, serial number, unique identifier (udi) # g3/4: corrected the following: pma/510(k)number.H4: corrected the following: device manufacture date.H6: corrected the following: health effect - clinical code, medical device problem code, component code, type of investigation.The reason for the revision reported cannot be confirmed from the information provided but may be the result of prosthesis wear and instability or due to inclusion of the polyethylene in the packaging recall.Potential contributions of user and patient-related considerations to the event could not be assessed as the devices were not available for evaluation and images and radiographs were not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIFIC DEVICE NOT REPORTED
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH INC
2320 nw 66th court
gainesville FL 32653
Manufacturer Contact
geoff gannon
2320 nw 66th court
gainesville, FL 32653
3523771140
MDR Report Key19242797
Report Number1038671-2024-01046
Device Sequence Number7495595
Product Code JWH
UDI-Device Identifier10885862001740
UDI-Public10885862001740
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K033883
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2023
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Consumer,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/19/2022
Device Catalogue Number02-012-35-2013
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/10/2024
Supplement Date Received by Manufacturer11/13/2024
Initial Report FDA Received Date05/03/2024
Supplement Report FDA Received Date11/15/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured03/24/2014
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-