|
See d4 expiration date, h4 and h11.The reason for this revision surgery, the agent reported "(patient fell and fractured pelvis.Cup was loose)".The actual length of in-vivo for the items listed is unknown as the original surgery date was not provided or could be established.This evaluation is limited in scope as the items associated with this investigation were not returned to djo surgical - austin for review.The surgery was completed as intended and without incident.If additional information regarding the reported event is submitted at a future date, this investigation will be re-evaluated.There was no information submitted with this complaint about any patient activities, accidents, or medical contraindications that may have contributed to the reported event.There were no findings during this evaluation that indicate the reported devices were defective.The surgeon performed this procedure to remedy the patient's condition.No further action is deemed necessary.Note: the previous d-ticket was not provided and a search of djo records produced no results, therefore, the items could not be verified.The lot numbers are incorrect, so changed to unknown.A review of the device history records (dhr) show that the reported components used in the previous surgery, when released for use, met design and manufacturing requirements.There was a nonconformance associated with the main part #940-02-54g, empowr acet system, cup, hemispherical, cluster hole, 54mm which documents that out of 24 parts lot, 3 parts were rejected and scrapped during dimensional and visual inspection.All other items in the lot were met with fit, form and function requirements.The devices were verified to have gone through an acceptable sterilization process and were within its expiration date at the time of the previous surgery.Customer complaint history of the reported devices showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to fracture and loosening after a fall.No information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.Agent has clearly mentioned that "patient fell" and there are multiple factors that may also contribute to an event that are outside the control of djo surgical such as poor bone density, inadequate soft tissue support, patient activities.
|