| Model Number |
VTICMO 13.2 |
| Medical Device Problem Code |
No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/02/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
The reporter indicated that a 13.2mm vticmo13.2 ,-16.0/3.0/107 (sphere/cylinder/axis), implantable collamer lens was implanted into the patient's right eye (od) on (b)(6) 2024.The patient experienced excessive vaulting.Intraoperatively the lens was removed and replaced with a shoter length lens and the problem resolved.The cause of the event was reported as unknown.
|
| |
|
Additional Manufacturer Narrative
|
|
Claim# (b)(4).
|
| |
|
Search Alerts/Recalls
|