| Catalog Number |
ASC4251-01 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2024
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an arthroscopy, the black alloy of the starvac 90 wand electrode came off in the joint when it was released; the detached particles were sucked in directly and did not remain in the joint.The electrode was triggered for approx.2-3 minutes without interruption in the joint under nacl.The procedure was completed using a s+n back up device.There was no surgical delay.No further complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference (b)(4) h10 h3, h6: a device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
H3, h6: the reported device was received for evaluation.Visual inspection observed the returned instrument shows no manufacturing abnormalities.The electrode tip has been heavily used but is intact.The black wand wrap is deformed at the distal end.Bio debris is present.Product was out of the original packaging.No packaging returned.A functional evaluation found the device displayed settings (1,7) when plugged in.The wand produced plasma and coagulation as expected.The resistance measured at 0.90 kilo ohms.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include 1) mechanical displacement of tissue through applied force or using the device as a lever to enlarge a surgical site or gain access to tissue.2) contacting the distal portion of cannula with the shaft when inserting and removing the wand.Based on this investigation, the need for corrective action is not indicated.
|
| |
|
Search Alerts/Recalls
|