• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHREX, INC. SUTR ANCH,FASTAK II; SCREW, FIXATION, BOLT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHREX, INC. SUTR ANCH,FASTAK II; SCREW, FIXATION, BOLT Back to Search Results
Model Number SUTR ANCH,FASTAK II
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/14/2022
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during an arthroscopy surgery the anchor could not be inserted.No part of the device broke off.According to the surgeon no harm for patient, operator or third party occurred.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
 
Additional Manufacturer Narrative
This 3500a record is submitted to comply with an fda 483 inspectional observation issued to arthrex inc on (b)(6) 2023.Arthrex has reassessed the reportability decisions made on historical complaint records using revised criterion.This 3500a document is a result of the reassessment.The complaint device was not received for evaluation.Based off the information provided, the most likely cause for the reported failure can be attributed to user error due to misaligned insertion and/or improper bone preparation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUTR ANCH,FASTAK II
Common Device Name
SCREW, FIXATION, BOLT
Manufacturer (Section D)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer (Section G)
ARTHREX, INC.
1370 creekside boulevard
naples FL 34108 1945
Manufacturer Contact
jared engle
8009337001
MDR Report Key19247590
Report Number1220246-2024-02800
Device Sequence Number12409657
Product Code HWC
UDI-Device Identifier00888867010376
UDI-Public00888867010376
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
K061863
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 05/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSUTR ANCH,FASTAK II
Device Catalogue NumberAR-1324HF-1
Device Lot Number14951921
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/15/2022
Initial Report FDA Received Date05/06/2024
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured04/19/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-