It was reported that, during a uka surgery, when adjusting the rm/ll tibia resection guide, the tip of the 3.5mm locking hex t-handle broke off at the front.All the broken pieces were removed with suction.Nothing remained in the patient.There was no surgical delay as consequence of this problem.
|
|
Additional information: h4 (manufacturing date), h6 (component code, type of investigation, investigation findings, investigation conclusions).Results of investigation: the associated device was returned and evaluated.The visual inspection revealed that the tip of the t-handle had been fractured off, showing visible signs of wear and use.A review of the production order did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history revealed similar events for the listed device over the previous 12 months, but no similar events for the batch based on the historical data, this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.The anticipated risk level is still adequate.A historical review concluded that there are no prior actions related to this product and event.At this time, we have no evidence to conclude that the product failed to meet any specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely potential factors that could contribute to the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|