One 7f, quadripolar, medium curl, livewire tc ablation catheter was received for evaluation.A short circuit was detected between electrode 1 and the thermistor wires.Dissection revealed that the flat wire insulation was torn and the insulation for the thermistor wires was abraded in the same location, consistent with the short circuit detected.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the torn insulation, abraded wires and short circuit remains unknown.
|